1.3 SPECIFICATIONS, G-M DETECTOR CHECK SOURCE,

MODEL 897A SERIES

Radionuclide

36

CI

Source Activity:

8 microcuries

Scale Reading (Approximate)
897A-210,

897A-211

897A-220,

897A-221

897A-230,

897A-231


10-150 mR/h, Typical
10-50 mR/h, Typical
10-50 mR/h, Typical

Voltage

12 – 15 Vdc

Encapsulation

Brass, high temperature soldered

1.4 SPECIFICATION, DETECTOR PREAMPLIFIER

(INTERGRAL)

Input Impedance

>100 k-ohms

Output Impedance

50 ohms

Output Pulse Polarity

Positive

Output Signal (50 ohm terminated)

+ 5 Vdc square-wave

Low Voltage*

+ 15 Vdc (Optional + 10 Vdc is Jumper Selectable)

Power Requirements

+ 15 Vdc @ 120 mA

Maximum Cable Length

2000 feet

Electronic Exposure Life

Approximately 10

5

rads

Discriminator Level

Adjustable from 0 to 1.5 Vdc (nominal value =1.00
Vdc)

Anti-Jam Level

Adjustable from 0 to + 3 Vdc

Anti-Jam Oscillator Frequency

Approximately 50 kHz

Pre-divide* Jumper

Selectable

Analog & Digital Monitor Configuration

Divide by 2, Square wave output

Optional Monitor Configuration

Raw pulse output

* Preamplifier low voltage and pre-divide are selected with jumpers. Refer to the preamplifier circuit

description for more information. Standard Configuration is +15 Vdc Divide by 2.