7 power up procedure – Fluke Biomedical 956A-201-M2 User Manual

Page 39

Advertising
background image

3-13

Anti-Jam Selection

Jumper JP6 is normally installed in position 1-2. Position 2-3 bypasses the anti-jam fuse for test purposes.

Detector Type

Jumper JP7 is set to position 2-3 for a G-M detector input. Position 1-2 is provided for use with a
scintillation detector. Removing the Jumper disables the anti-jam circuit for test purposes.

Table 3-2. Model 956A-201-M2 Option Jumpers

Jumper Function

Position

Operation

JP1 Microprocessor

Reset Out



IN

Normal Operation (factory) Momentary connect to
reset microprocessor without cycling AC power
Not Applicable

JP2 PROM

Type

1-2
2-3

PROM 27128
PROM 27256 (factory)

JP3-1/JP3-
2

Statistical Accuracy (Note: this
accuracy refers only to the
statistical interpretation of
detector counts, and not
detector accuracy)

Out/Out
Out/In
In/Out
In/IN

2% Accuracy – 9604 Target Counts, Not
implemented
5% Accuracy – 1537 Target Counts, Not
implemented
10% Accuracy – 384 Target Counts, Not
implemented
Fixed one second display update of 60 one
second values (factory)

JP3-3

Alarm Acknowledge

In
Out

Manual Acknowledge (factory)
Automatic

JP3-4 Fail

Alarm

In
Out

Enable No Counts Fail Alarm (factory)
Inhibit No Counts Fail Alarm

JP3-5

Check Source Alarm

In
Out

Alarm Inhibited (factory)
Alarm Enabled

JP4

Input Pulse Selection

1-2
2-3

Negative Input Pulse (Other Detectors)
Positive Input Pulse (G-M Detectors) (factory)

JP5

Shield Polarity Selection

1-2
2-3

Shield for Negative Pulse
Shield for Positive Pulse (G-M) (factory)

JP-6

Anti-Jam Fuse Selection

1-2
2-3

Enable for Normal Operation (factory)
Anti-Jam Circuit fuse bypassed (testing)

JP7

Detector Type for Anti-Jam
Circuit Timing

1-2
2-3
Out

Scintillation
G-M Tube (factory)
Anti-Jam Circuit Disabled (for testing only)

3.7 POWER UP PROCEDURE

To place the system in operation, the following steps should be performed:

1. Verify that the UDR configuration jumpers are properly selected. Refer to Section 3.6 "Configuration

Functions" for addition information if required.

2. Verify that the UDR Function switch, located on the front right section of the UDR main board, is in a

position other than 8.

3. Verify that the detector and UDR are properly connected.

4. Apply power to the system. If all connections are good, the UDR will power up and will begin to

monitor the ambient radiation level.

Advertising
This manual is related to the following products: