3-15

Table 3-4. Function Switch Entry Parameters

Switch
Position

Setpoint Units

Detector

897A-21X

Detector

897A-22X

Detector

897A-23X

Default

Value

0

High Alarm Limit

mR/h

**

**

**

1.00E3

1

Warn Alarm Limit

mR/h

**

**

**

1.00E1

2 Resolving

Time

(Dead Time)

minutes *

*

* 0.00E0

3

Analog Full Scale
Value

mR/h 1.00E3

1.00E4

1.00E5

1.00E4

4 Overrange

Limit

mR/h

1.00E3

1.00E4

1.00E5

1.00E4

5 Conversion

Constant

mRh/CPM

* * *

1.00E-2

6 Not

Used N/A

E0007

E0007

E0007

E0007

7 Analog

Low

Scale

Value

mR/h 1.00E-2

1.00E-1

1.00E0

1.00E-1

8 Calibration

Mode/Time

seconds **

**

** 6.00E1

9

Underrange Limit

mR/h

1.00E-2

1.00E-1

1.00E0

1.00E-1

A through F

Not Used

-

E0007

E0007

E0007

E0007

* Detector unique, provided with the detector/detector calibration sheet.
** User Selected (detector range dependent).

To enter or display a parameter, the FUNCTION switch is set to the desired position and the following
steps are performed:

In the event it is necessary to re-enter all of the default setpoints,
the UDR EPROM memory may be reinitialized by turning the
UDR power off, then depressing the ENTER pushbutton at the
same time the power on switch is depressed. Releasing the
ENTER pushbutton and momentarily depressing the ENTER
pushbutton again to exit the setpoint entry mode and return to
normal operation.

1. Press the ENTER pushbutton. The current value of the parameter, selected by the FUNCTION switch

will be displayed on the front panel. The leftmost digit of the value will be flashing. The value is
displayed in the format shown below.

M.MMEN

This format expresses the form of M.MMEN, where M is a whole number from 0 to 9, E designates a
positive exponent, and N represents the exponent value from 0 to 9. A negative exponent is
expressed as a minus (-) character in the E position. Pressing the Enter button again enters the
value currently displayed into non-volatile setpoint memory.

2. The value of the flashing digit can be incremented by pressing the VALUE pushbutton.

3. The flashing digit can be moved one place to the right by pressing the DIGIT pushbutton.

4. When the desired value has been entered into the display, pressing the ENTER pushbutton causes

the displayed value to be entered into non-volatile memory for permanent storage. The new
parameter value is now effective.

NOTE