Fluke Biomedical 956A-201-M2 User Manual

Page 22

Advertising
background image

2-4

Table 2-4. Connector P1 – Input/Output, 956A-201-M2

Pin Signal

Internal

Connection

1

120 Vac Line

Relay Board K1-A/V2

2 120

Vac

Neutral

P2-13

3 120

Vac

Ground

P2-11

4 Spare

Relay

Board

K1-D

5 Spare

Relay

Board

K1-F

6 Spare

Relay

Board

K1-E

7

Fail Relay, Common

Relay Board K3-A

8

Fail Relay, n.o.

Relay Board K3-B

9

Fail Relay, n.c.

Relay Board K3-C

10

Fail Relay, Common

Relay Board K3-D

11

Fail Relay, n.o.

Relay Board K3-F

12

Fail Relay, n.c.

Relay Board K3-E

13

Warn Relay, Common

Relay Board K4-A

14

Warn Relay, n.o.

Relay Board K4-B

15

Warn Relay, n.c.

Relay Board K4-C

16

Warn Relay, Common

Relay Board K4-D

17

Warn Relay, n.o.

Relay Board K4-F

18

Warn Relay, n.c.

Relay Board K4-E

19

Alarm Relay, Common

Relay Board K5-A

20

Alarm Relay, n.o.

Relay Board K5-B

21

Alarm Relay, n.c.

Relay Board K5-C

22 Spare

Not

Used

23 Spare

Not

Used

24 Spare

Not

Used

25

+ Remote Acknowledge

Main Circuit Board J9-1

26

- Remote Acknowledge

Main Circuit Board J9-2

27 - 37

Spare

Not Used

n.o. = normally open, n.c. = normally closed, Relays shown in shelf, or de-energized state.

Table 2-5. Connector P2 – Detector Connector, 956A-201-M2

Pin Signal

Internal

Connection

1

Detector + 15 Vdc Supply

Power Supply + 15 Vdc

2

Not Used

Power Supply – 15 Vdc

3

Supply Ground

Power Supply Ground

4

+ 15 V C/S On

Relay K2

5

- 15 V C/S Off

Relay K2

6 Not

Used

-

7 Not

Used

-

8 Not

Used

-

9

DC Ground

Main Circuit Board J5-9

10

0 – 10 Vdc, Remote Meter

Main Circuit Board J5-10

11

AC Ground

P1-3 AC Ground

12

Fail, N.C (Switched, 120 Vac, Line)

Relay Board K1-C (L)

13

120 Vac (Neutral)

P1-2, 120 Vac (N)

14

Alarm N.C (Switched, 120 Vac, Line) Relay Board K5-E (L)

0 – 10 Vdc selected for Remote Meter use on P2 (10, 11) not P6 (5, 6).

Advertising
This manual is related to the following products: