Table 2-12. Model 956A-201-M2 Option Jumpers

Jumper Function

Position Operation

JP1 Microprocessor

Reset

Out


In

Normal Operation (factory) Momentary
connect to reset microprocessor without
cycling AC power
Not Applicable

JP2 PROM

Type

1-2
2-3

PROM 27128
PROM 27256 (factory)

JP3-1/JP3-

2

Statistical Accuracy (Note: This
accuracy refers only to the
statistical interpretation of
detector counts, and not
detector accuracy)

Out/Out
Out/In
In/Out
In/In

2% Accuracy – Not Implemented
5% Accuracy – Not Implemented
10% Accuracy – Not Implemented
Fixed one second display update (factory

JP3-3 Alarm

Acknowledge

In
Out

Manual Acknowledge (factory)
Automatic

JP3-4 Fail

Alarm

In
Out

Enable No Counts Fail Alarm (factory)
Inhibit No Counts Fail Alarm

JP3-5

Check Source Alarm

In
Out

Alarm Inhibited (factory)
Alarm Enabled

JP4

Shield Polarity Selection

1-2
2-3

Negative Input Pulse (Other Detectors)
Positive Input Pulse (G-M Detectors)
(factory)

JP5

Shield Polarity Selection

1-2
2-3

Shield for Negative Pulse
Shield for Positive Pulse (G-M) (factory)

JP-6

Anti-Jam Fuse Selection

1-2
2-3

Enable for Normal Operation (factory)
Anti-Jam Circuit fuse bypassed (testing)

JP7

Detector Type for Anti-Jam
Circuit Timing

1-2
2-3
Out

Scintillation
G-M Tube (factory)
Anti-Jam Circuit Disabled (for testing only)