3-37

Table 3-17. Option Jumper Select Read Functions

Data Bit

Function

D0

Serial data from E

2

PROM

D1

Anti-Jam

D2

(Not

used)

D3

JP3-5

(IN) Inhibit alarms during check source operation

(OUT) Alarms active during check source operation

D4

JP3-4

(IN) Fails in five minutes with no counts

(OUT) Does not fail

D5 JP3-3

(OUT) Auto alarm acknowledge, after counts return to
normal

(IN) Manual Reset (factory setting)

D6

JP3-2

IN/IN, Fixed 1 second update

D7

JP3-1

IN/OUT, 384 Target Counts, 10% accuracy
OUT/IN, 1537 Target Counts, 5% accuracy
OUT/OUT, 9604 Target Counts, 2% accuracy

JP3 jumper IN MPU reads a low (0)

JP3 jumper OUT MPU reads a high (1)

Switch Inputs

Octal inverting buffer U62 functions as an interface to provide the status of the front panel control switches
to the internal data bus and MPU. U62 responds to address 4000 and / SWITCH INTERFACE. The
switch input bit assignments are listed in table 3-18.

Table 3-18. Switch Input Bit Assignments (U62)

Data Bit

Function

D0 Alarm

Acknowledge

D1 Check

Source

D2

(Spare, Input 2)

D3

(Spare, Input 1)

D4

Rate-of-Rise (not used)

D5 Warn

Alarm

D6 High

Alarm

D7

High Voltage (not used)


When a switch is pressed, the appropriate input to U62 is pulled low. When U62 is read by the MPU, a
high (1) is available on the data bus. When no switches have been pressed, all outputs (U62) will be low
when read. U63A and B are D Type Flip-Flops that latch switch data from the check source and alarm
acknowledge switches. The MPU controls the clear switch latch signal to reset U63A and B. The circuit
comprised by S1, D1, and R11 is a remote alarm acknowledge. R11 is selected to allow a 20 mA signal to
flow through the S1 infrared diode when a given voltage is present on the J9 remote acknowledge input.
When this voltage is present, the S1 infrared diode is forward biased, causing the S1 phototransistor to
conduct. This effectively forces a low (0) to U63-10, setting the alarm acknowledge bit.

Option Board Bus

The option board bus is available on connector J3. All address, data, and control signals are provided to
allow various digital/analog circuit boards to directly interface to the main circuit board. Decoded signals
for asynchronous communications interface adapter option, and the general-purpose interface bus option
are available on the option board interface connector.