Fluke Biomedical 956A-201-M2 User Manual

Page 23

Advertising
background image

2-5

Table 2-6. Connector P3 – Power Input

Pin Signal

Internal

Connection

1

120 Vac, Line

Line fuse (F2)

2

120 Vac, Neutral

Power Supply, 120 Vac (n)

3 Safety

Ground Chassis


Table 2-7. Connector P4 – Detector High Voltage

Pin Signal

Internal

Connection

MHV

Detector High Voltage

Direct


Table 2-8. Connector P5 – Detector Signal Input

Pin Signal

Internal

Connection

BNC Detector

Signal

Direct

Table 2-9. Connector P6 – Auxiliary I/O, 956A-201-M2

Pin Signal

1

4 – 20 mA Output #1 (+)

2

4 – 20 mA Ground (-)

3

4 – 20 mA Output #2 (+)

4

4 – 20 mA Ground (-)

5

0 – 10 V* (+)

6 Ground

(-)

7

Analog Output Option (+), 0-10 mV

8 Ground

(-)

9 through 24

Not used

*0 – 10 Vdc selected for Remote Meter use on P2 (10, 11) and is not available on P6 (5, 6).

Table 2-10. Connector P7 – RS232C Option

Pin Signal

1 Protective

Ground

2 Transmit

Data

3 Receive

Data

4

Request to Send (RTS)

5

Clear to Send (CTS)

6

Data Set Ready (DSR)

7 Signal

Ground

8

Receive Line Signal Detect (DCD)

20

Data Terminal Ready


Table 2-11. 897A Detector Connection

Pin Signal

K Protective

Ground

H Check

Source

J Check

Source

E High

Voltage

MTG Screw

Chassis Ground

C

+ Voltage In

G Signal

Output