2 specifications, g-m detector, model 897a series – Fluke Biomedical 956A-201-M2 User Manual

Page 7

Advertising
background image

1-3

1.2 SPECIFICATIONS, G-M DETECTOR, MODEL 897A SERIES

Dimensions (Approximate

7 1/8 x 3 inches (18.1 x 7.6 cm)

Weight (Approximate)

2.5 lbs.

Housing Material
897A-210, 897A-220, 897A-230
897A-211, 897A-221, 897A-231


Aluminum, weatherproof
Stainless Steel, weatherproof

Fill Gas:
897A-210, 897A-211
897A-220, 897A-221
897A-230, 897A-231


Neon/Argon/Halogen
Helium/Neon/Halogen
Helium/Neon/Halogen

G-M Tube Wall Thickness:
897A-210, 897A-211
897A-220, 897A-221
897A-230, 897A-231


40 mg/cm

2

80 mg/cm

2

80 mg/cm

2

Mounting Wall

mount

Mating Connectors

897A-2x0:
92-7005-17A, 12 Pin Female
92-7005-12A,

Bushing

92-7005-9A,

Clamp


897A-2x1 (stainless steel):
92-7005-15A, 12 Pin Female (stainless steel)
92-7005-12A,

Bushing

92-7005-13A,

Bushing

Required Cable

P/N 50-100 or substitute with two coaxial
conductors, two twisted pairs, and overall shield.

External Pressure Limit

30 psig

Storage Temperature

-10° to 122°F (-23° to 50°C)

Operating Temperature

-10° to 122°F (-23° to 50°C)

Relative Humidity

0 to 95% non-condensing

Operating Voltage

500 – 650 Vdc (supplied by digital ratemeter)

Plateau Length

100 – 150 Vdc

Plateau Slope:
897A-210,

897A-211

897A-220,

897A-221

897A-230,

897A-231


0.1%/V
0.2%V
0.3%V

Dead Time (Approximate)
897A-210,

897A-211

897A-220,

897A-221

897A-230,

897A-231


45 microseconds
28 microseconds
20 microseconds

Detector Element Life

Exceeds 1000 hours at full-scale

Energy Dependence of Reading

+/- 15% from 80 keV to 1.5 MeV

Detector Accuracy

±

20% of actual dose for Cs-137

Radiation Detected

Gamma rays, X-rays

Advertising