Introduction and Specifications

Introduction

1

1-3

Introduction

The Fluke Biomedical BP Pump 2 Non-Invasive Blood Pressure Simulator and
Tester, hereafter known as the “Tester”, is a multi-purpose test instrument for
use with oscillometric Non-Invasive Blood Pressure Monitors (NIBPMs). The
Tester provides dynamic blood pressure simulations, static calibration,
automated leak testing, and pressure relief valve testing. The following
models are available:

BP Pump 2L (Basic Model)

BP Pump 2M (High-Accuracy Model)

The Tester allows you to verify the performance claims of different blood
pressure monitors. You can quickly recall the fixed onboard simulations or
define your own. With its internal pump, the Tester can generate pressures up
to 400 mmHg (53.3 kPa) for leak testing, pressure sourcing, and relief valve
testing.

In addition, you can define auto sequences that automate the sequencing of
tests and NIBP simulations and provide an optional printed report.

Key Features

Key features of the Tester include:

Pressure leak testing on cuff, tubing, and connections

Relief valve testing on the patient monitor

Pressure gauge measurements

Pressure source capability

NIBP simulations including adult, neonate, arrhythmias, and respiratory

artifacts

Auto sequences with optional reports

Internal Adult and Neonatal Cuff simulation

Tester capabilities can be extended with optional accessories that allow:

ECG synchronization with non-invasive output

External wrist cuff simulations