Simulations – Fluke Biomedical BP Pump 2 User Manual

Page 49

Advertising
background image

Operation

Simulations

3

3-13

Note

The

<START>

soft key is not applicable to the pressure gauge test and

does not display.

3. Apply pressure to the pressure port and read the displayed pressure in the

upper right corner of the screen.

Note

You can use a squeeze bulb or syringe to apply adequate pressure.

4. Press

HOME

to return to the

Main

menu.

Simulations

The following sections describe various simulations that the Tester accomplishes.
For all of the following tests, it is important to select the correct cuff.

To perform simulations, take the following actions:

1. Power on the NIBP unit.

Note

Either disable the alarm or be prepared to silence it when it sounds.

2. Use the soft keys, beginning from the

Main

menu:

Fl

u

ke Biomedical BP B

u

mp 2

SETUP

PRESSURE

TESTS

AUTO

SEQUE

N

CE

PERFORM

SIMULATIO

N

fas31.eps

3. Press the

PERFORM SIMULATION

soft key to access the

Select

Simulation Type

menu. From this menu, you can access the simulations

described below:

Advertising