Relief valve test, Pressure source – Fluke Biomedical BP Pump 2 User Manual

Page 62

Advertising
background image

BP Pump 2

Operators Manual

3-26

2. If the Leak Rate is unsatisfactory, repair the possible cause of the leak and

press

START

again to repeat the test.

3. Press the

NEXT>

key to advance to the next test. Only the leak rate displayed

when the

NEXT>

key is pressed appears on the printout.

Note

The

NEXT>

key will not function until the Leak Test has been performed

at least once.

Relief Valve Test

This test increases the pressure in the pneumatic system until the relief valve on
the NIBP monitor opens or until the Setpoint is reached, whichever occurs first.

1. Press the

START

soft key to initiate this test. It is recommended that this test

be repeated multiple times.

2. Press the

NEXT>

key to advance to the next test.

Note

The

NEXT>

key will not function until the Relief Valve Test has been

performed at least once.

Pressure Source

This test causes the system to rise to the Setpoint pressure, which was defined
when the auto sequence was edited.

1. Press the

START

soft key to initiate this test. The system then begins to

pressurize until a stable pressure reading at or near the Setpoint is reached.

2. Record the monitor reading on the data sheet (if generated).

3. Press the

NEXT>

key to advance to the next test.

Note

The

NEXT>

key does not function until the Pressure Source test has been

performed.

Advertising