Remote operation, Rs-232 settings, Ansur software control – Fluke Biomedical BP Pump 2 User Manual

Page 65

Advertising
background image

Operation

Remote

Operation

3

3-29

Remote Operation

To prepare for remote operation, do the following:

1. Power down the Tester and connect the Tester to your PC, using a standard

bidirectional RS-232 cable.

2. Turn the Tester on: the operating system recognizes the Tester as new

hardware.

You are now ready to operate the Tester remotely.

RS-232 Settings

The Tester serial port parameters are fixed at the following settings:

Baud Rate: 9600

Parity: None

Data Bits: 8

Stop Bit:

1

Ansur Software Control

Ansur test automation systems allow a solutions-based approach to complete
testing of the medical device under test (DUT). Ansur helps you create standard
work using the test template/sequence (which is based on your written test
procedure) and integrates all test results into a single test report that can be printed
or archived. Ansur manages your test procedures by allowing both manual and
visual automated test sequences.

The software works hand-in-hand with Fluke Biomedical testers and simulators,
creating a seamless integration for:

Visual inspections

Preventive maintenance

Work procedures

Performance tests

Safety tests

Advertising