Respiratory artifacts – Fluke Biomedical BP Pump 2 User Manual

Page 54

Advertising
background image

BP Pump 2

Operators Manual

3-18

Respiratory Artifacts

The Respiratory Artifact exhibits a beat-to-beat variation in the blood pressure
caused by intra-thoracic pressure. Changes in the intra-thoracic pressure affect
filling of the ventricles during diastole. This in turn affects the stroke volume of
the heart. A large stroke develops a higher systolic pressure than a small stroke.

To perform Respiratory Artifact simulations, do the following:

1. Access these simulations by pressing the

7 RESPIRATORY ARTIFACT

key. The

Respiratory Artifact

screen appears.

Respiratory Artifact
Spont Breath # 1
C

u

ff: External

OPTIO

N

S

CUFF

ready ....
13

8

/

65 (

8

9)

104 BPM

fas11.eps

2. Press the

OPTIONS

soft key to scroll through the simulation choices.

Parameters for Respiratory Artifact simulations are shown in Table 3-5.

Table 3-5. Respiratory Artifact Simulations

Artifact Type

Blood

Pressure

(mmHg) (MAP)

Heart Rate

(bpm)

Pulse

volume (cc)

Spontaneous Breathing #1

138/65 (89)

104

Varies

Spontaneous Breathing #2

149/65 (93)

105

Varies

Spontaneous Breathing #3

112/47 (68)

86

Varies

Controlled Ventilation

132/44 (73)

98

Varies

3. Press the

CUFF

soft key to select Internal Adult or External cuff.

4. Press

Start

on the NIBP unit.

5. Compare values on the NIBP unit with those on the Tester.

6. Press

HOME

to return to the

Main

menu.