Bp pump 2, Operators manual, Figure 3-8. sample data sheet printout – Fluke Biomedical BP Pump 2 User Manual

Page 64

Advertising
background image

BP Pump 2

Operators Manual

3-28

Fluke Biomedical

Tester Serial #:123456

Date:________

Time:_______

Serial #: ___________________

Control #: __________________

Mfg: ________________________

Model: ______________________

Location: ___________________

Technician: _________________

Work Order: _________________

Procedure ID: _______________

Gauge: 181 mmHg

Leak Test

Leak Rate:

5 mmHg/min

Start Pressure

203 mmHg

End Pressure:

198 mmHg

Relief Valve Test

Peak Pressure:

No Relief Detected

418 mmHg

Pressure Source

Actual:

Source:

_________

186 mmHg

Actual

___/___ ( ) ___BPM

Preset #1

120/80 80BPM 0.7cc

Actual

___/___ ( ) ___BPM

Preset #2

150/100 80BPM 0.7cc

Actual

___/___ ( ) ___BPM

Preset #3

200/150 80BPM 0.7cc

Actual

___/___ ( ) ___BPM

Preset #4

255/195 80BPM 0.7cc

Actual

___/___ ( ) ___BPM

Preset #5

60/30 80BPM 0.7cc

Figure 3-8. Sample Data Sheet Printout

Advertising