Running auto sequences – Fluke Biomedical BP Pump 2 User Manual

Page 60

Advertising
background image

BP Pump 2

Operators Manual

3-24

Auto Sequence #1 Definition

Gauge Enabled

Leak Setpoint:

200 mmHg

Relief Pressure:

380 mmHg

Source Pressure:

200 mmHg

Cuff: External

Simulation 1:

Preset #1

120/80

80BPM 0.7cc

Simulation 2:

Preset #2

150/100

80BPM 0.7cc

Simulation 3:

Preset #3

200/150

80BPM 0.7cc

Simulation 4:

Preset #4

255/195

80BPM 0.7cc

Simulation 5:

Preset #5

60/30

80BPM 0.7cc

Figure 3-7. Sample Auto Sequence Test Printout

Running Auto Sequences

Use the following steps to run an auto sequence.

1. Press the

AUTO SEQUENCE

soft key to access the

Auto Sequence

menu.

A

u

to Se

qu

ence

EDIT

PRI

N

T

RU

N

fas36.eps

2. Press the

RUN

soft key. The

Select Auto Sequence (1-9)

screen appears.