BP Pump 2
Operators Manual
3-24
Auto Sequence #1 Definition
Gauge Enabled
Leak Setpoint:
200 mmHg
Relief Pressure:
380 mmHg
Source Pressure:
200 mmHg
Cuff: External
Simulation 1:
Preset #1
120/80
80BPM 0.7cc
Simulation 2:
Preset #2
150/100
80BPM 0.7cc
Simulation 3:
Preset #3
200/150
80BPM 0.7cc
Simulation 4:
Preset #4
255/195
80BPM 0.7cc
Simulation 5:
Preset #5
60/30
80BPM 0.7cc
Figure 3-7. Sample Auto Sequence Test Printout
Running Auto Sequences
Use the following steps to run an auto sequence.
1. Press the
AUTO SEQUENCE
soft key to access the
Auto Sequence
menu.
A
u
to Se
qu
ence
EDIT
PRI
N
T
RU
N
fas36.eps
2. Press the
RUN
soft key. The
Select Auto Sequence (1-9)
screen appears.