“The mean difference of the paired measurements of the test system and the

comparison system shall be ± 5 mmHg or less with a standard deviation of

8 mmHg or less.”

This means that variations in individual readings of 5, 6, or even 10 mmHg are

quite normal and do not indicate that either the DINAMAP or the Tester are

malfunctioning. Some monitors are more repeatable than others, and

repeatability is one measure of the overall quality of the monitor.

Q: BP results vary using the same preset pressure.

“I checked another NIBP monitor using the same preset simulated blood

pressure of 120/80 (93) with a pulse rate of 80 bpm. This time I got the

following results:

Trial #

Systolic

Mean

Diastolic

Pulse Rate

1 120 89 71 80

2 120 87 73 80

3 121 91 72 80

Why does this monitor show such low Diastolic pressures?”

A: Monitors using different references – Ascultatory vs. Invasive data.

Neither the monitor nor the Tester is broken or giving incorrect readings.

Some monitors were designed to give readings close to those obtained by the

Ascultatory method of blood pressure determination. Other monitors have been

designed to agree with Invasive Blood Pressure readings. It is well known that

Invasive and Ascultatory BP readings on the same subject can be quite

different. Therefore it is not surprising that automated Oscillometric NIBP

monitors using Invasive readings as a reference would give different readings

than a monitor based on Ascultatory readings.

Cuff Issues

Q: Why is an Internal Cuff used?

“Why does the Tester use an Internal Simulated Cuff? Wouldn’t it be better to

include a real cuff in the measurement?”

A: Internal Cuff produces accurate and repeatable simulations.

The Tester uses an internal cuff to help ensure accurate and repeatable

simulations over time. The internal “cuff” is a 310-ml fixed volume that has

compliance very nearly equal to a normal adult cuff when used at typical adult