Neonate – Fluke Biomedical BP Pump 2 User Manual

Page 55

Advertising
background image

Operation

Simulations

3

3-19

Neonate

The Neonate simulations are provided to test the ability of the NIBP monitors to
detect blood pressure on neonatal patients.

To perform Neonate simulations, do the following:

1. Access these simulations by pressing the

8 NEONATE

key. The

Neonate

Simulation

screen appears.

N

eonate Sim

u

lation

N

eonate # 1

C

u

ff: Internal

N

eonate

OPTIO

N

S

CUFF

ready ....
35/

15(22)

120 BPM 0.11 cc

fas12.eps

2. Press the

OPTIONS

soft key to scroll through the simulation choices.

Parameters for Neonate simulations are shown in Table 3-6.

Table 3-6. Neonate Simulations

Simulation

Number

Blood Pressure

(mmHg) (MAP)

Heart Rate (bpm)

Pulse volume

(cc)

1 35/15

(22)

120 0.11

2 60/30

(40)

120 0.10

3 80/50

(60)

120 0.10

4 100/70

(80)

120

0.10

3. Press the

CUFF

soft key to select External or Internal Neonate.

4. Press

Start

on the NIBP unit.

5. Compare values on the NIBP unit with those on the Tester.

6. Press

HOME

to return to the

Main

menu.