Arrhythmias – Fluke Biomedical BP Pump 2 User Manual

Page 52

Advertising
background image

BP Pump 2

Operators Manual

3-16

Patient Condition
Healthy
C

u

ff: External

OPTIO

N

S

CUFF

ready ....
120/

8

0 (93)

75 BPM

0.6

8

cc

fas08.eps

2. Press the

OPTIONS

soft key to scroll through the simulation choices.

Parameters for Patient Condition simulations are shown in Table 3-3.

Table 3-3. Patient Condition Simulations

Patient Condition

Blood Pressure

(mmHg) (MAP)

Heart Rate

(bpm)

Pulse volume

(cc)

Healthy Heart

120/80 (93)

75

0.68

Weak Pulse

110/80 (90)

95

0.50

Mild Exercise #1

140/90 (106)

120

1.00

Strenuous Exercise #2

140/90 (106)

162

1.40

Obese Subject

120/80 (93)

90

0.50

Geriatric Subject

150/110 (123)

95

0.40

Tachycardia 120/105

(110)

130

0.40

Bradycardia 120/60

(80)

45

1.10

3. Press the

CUFF

soft key to select Internal Adult or External cuff.

4. Press

Start

on the NIBP unit.

5. Compare values on the NIBP unit with those on the Tester.

6. Press

HOME

to return to the

Main

menu.

Arrhythmias

These waveforms cause erratic readings on some NIBPMs. The blood pressure
determination strongly depends on exactly what is happening with the subject's
blood pressure when the cuff pressure is at a particular level. Some NIBPMs
pause until they detect two or more equivalent beats. The pattern of step deflations
and the measured blood pressure depend on which beats occur during each step of
the cuff pressure.