User-defined, Auto sequences – Fluke Biomedical BP Pump 2 User Manual

Page 57

Advertising
background image

Operation

Auto

Sequences

3

3-21

User-Defined

User-Defined simulations and how to create them are described in “Setting Up the
Tester.”

The following describes how to run such a simulation.

1. Select from any of the previously-defined simulations by pressing the

0

USER DEFINED

key. The

User Defined

screen appears.

User Defined
Sim

u

lation 1:

C

u

ff: External

OPTIO

N

S

CUFF

EDIT

ready ....
120/

8

0

75 BPM

0.70 cc

fas15.eps

2. Press the

OPTIONS

soft key to scroll through the simulation choices.

3. Press the

CUFF

soft key to select

External

,

Internal Adult

, or

Internal

Neonate

cuff.

Note

It is important to select the correct cuff.

4. Press

Start

on the NIBP unit.

5. Compare values on the NIBP unit with those on the Tester.

6. Press

HOME

to return to the

Main

menu.

Auto Sequences

It is possible to create up to nine customized auto sequences. An auto sequence
contains all four pressure tests and five simulations. The operator can disable any
of these tests or simulations. A printout of the auto sequence result can also be
enabled. The operator must make a cuff selection; this determines what NIBP
simulations are displayed for that selection.