User-defined simulations – Fluke Biomedical BP Pump 2 User Manual

Page 29

Advertising
background image

Setup, Maintenance, and Support

Setting up the Tester

2

2-5

User-Defined Simulations

The Tester supports up to nine user-definable blood pressure simulations
configured through the

Setup

menu.

To access and modify a user-defined simulation definition, take the following
actions:

1. Use the soft keys to follow this menu path:

SETUP

C

USER DEF SIMULATION

C

Number Key (1-9)

C

ENT

2. Alternatively, you can access the definition from the number keys:

0 USER DEFINED key

C

OPTIONS (to scroll through the

numbers)

C

EDIT

With either method, the

User Defined Simulation

screen appears:

User Defined Sim

u

lation 1

Sys: 120 Dia:

8

0

P

u

lse: 0.70cc

HR: 75

<- FIELD

FIELD ->

fas23.eps

This screen shows the number of the simulation, along with the following
parameters available for configuration. Their valid ranges are as follows:

Sys

: Systolic, 20 – 250 mmHg

Dia

: Diastolic, 10 – 200 mmHg

Pulse

: Pulse Volume, 0.1 cc – 2.4 cc in increments of 0.1 cc

HR

: Heart Rate, 30 – 250 bpm

The systolic and diastolic settings are interdependent. The diastolic must
always be below the systolic. The pulse volume and the heart rate are also
interdependent. The maximum pulse volume cannot be achieved at the
maximum heart rate.

3. Press the soft key

<- FIELD

or

FIELD ->

to scroll through options,

making changes to each field by entering numbers from the number keys.

4. Pres

ENT

to return to the

SETUP

menu.

5. If desired, press the

HOME

button to return to the

Main

menu.