Pressure leak test – Fluke Biomedical BP Pump 2 User Manual
Page 44
BP Pump 2
Operators Manual
3-8
All tests begin from the
Main
menu:
Fl
u
ke Biomedical BP B
u
mp 2
SETUP
PRESSURE
TESTS
AUTO
SEQUE
N
CE
PERFORM
SIMULATIO
N
fas31.eps
After pressing the
PRESSURE TESTS
soft key to display the
Pressure Tests
menu, you can access the pressure tests described below:
PRESSURE
RELIEF
STATIC
PRESSURE
PRESSURE
LEAK TEST
fas35.eps
Pressure Leak Test
The Pressure Leak Test pressurizes a pneumatic system to an operator-defined
target pressure (labeled Setpoint) up to 400 mmHg (53.3 kPa) and then measures
the loss of pressure over time.
To assess leakage, take the following actions:
1. Define the Setpoint by using the soft keys to follow the menu path shown
below.
PRESSURE TESTS
C
PRESSURE LEAK TEST
C
SETUP
The
Leak Test (Setpoint)
screen appears.
Leak Test
Setpoint: 250
fas41.eps
2. Adjust the Setpoint by entering numbers with the number keys
.
The cursor
moves to the next position after a number is entered