All tests begin from the

Main

menu:

Fl

u

ke Biomedical BP B

u

mp 2

SETUP

PRESSURE

TESTS

AUTO

SEQUE

N

CE

PERFORM

SIMULATIO

N

fas31.eps

After pressing the

PRESSURE TESTS

soft key to display the

Pressure Tests

menu, you can access the pressure tests described below:

PRESSURE

RELIEF

STATIC

PRESSURE

PRESSURE

LEAK TEST

fas35.eps

Pressure Leak Test

The Pressure Leak Test pressurizes a pneumatic system to an operator-defined
target pressure (labeled Setpoint) up to 400 mmHg (53.3 kPa) and then measures
the loss of pressure over time.

To assess leakage, take the following actions:

1. Define the Setpoint by using the soft keys to follow the menu path shown

below.

PRESSURE TESTS

C

PRESSURE LEAK TEST

C

SETUP

The

Leak Test (Setpoint)

screen appears.

Leak Test

Setpoint: 250

fas41.eps

2. Adjust the Setpoint by entering numbers with the number keys

.

The cursor

moves to the next position after a number is entered