Patient conditions – Fluke Biomedical BP Pump 2 User Manual

Page 51

Advertising
background image

Operation

Simulations

3

3-15

Standard BP
Preset # 1
C

u

ff: External

OPTIO

N

S

CUFF

ready ....
120/

8

0 (93)

8

0 BPM

0.6

8

cc

fas07.eps

2. Press the

OPTIONS

soft key to scroll through the simulation choices.

Parameters for Standard BP simulations are shown in Table 3-2.

Table 3-2. Standard Blood Pressure Simulations

Simulation

Number

Blood Pressure

(mmHg) (MAP)

Heart Rate (bpm)

Pulse volume

(cc)

1 120/80

(93) 80

0.68

2 150/100

(116) 80

0.65

3 200/150

(166) 80

0.60

4 255/195

(215) 80

0.55

5 60/30

(40) 80

0.75

6 80/50

(60) 80

0.71

7 100/65

(76) 80

0.69

3. Press the

CUFF

soft key to select Internal Adult or External cuff.

4. Press

Start

on the NIBP unit.

5. Compare values on the NIBP unit with those on the Tester.

6. Press

HOME

to return to the

Main

menu.

Patient Conditions

The Patient Condition simulations are intended to provide some basic patient
variations.

To perform Patient Conditions simulations, do the following:

1. Access these simulations by pressing the

5 PATIENT CONDITIONS

key.

The

Patient Condition

screen appears.

Advertising