Fluke Biomedical BP Pump 2 User Manual

Page 18

Advertising
background image

BP Pump 2

Operators Manual

1-6

Always connect the system power cord directly to a
three-prong receptacle with a functional ground. Never
use a two-prong plug adapter to connect primary
power to the Tester, thereby disconnecting the utility
ground.

Disconnect the Tester from the power source before
changing the supply voltage. The Tester operates at a
range of 100 to 240 volts.

W

Caution

To avoid

damage to the Tester or adverse affects on its

performance, follow these guidelines

:

Allow only qualified technical personnel to service the
Tester.

Do not expose the system to temperature extremes.
Ambient temperatures should remain between 15° C
and 40° C. System performance may be adversely
affected if temperatures fluctuate above or below this
range.

Clean the Tester only by gently wiping down with a
clean, lint-free cloth dampened with a mild detergent
solution. Do not immerse the unit.

Do not apply pressures greater than 400 mmHg (53
kPa) to the pressure port.