Editing Auto Sequences

Use the following steps to edit an auto sequence:

1. Press the

AUTO SEQUENCE

soft key to access the

Auto Sequence

menu.

A

u

to Se

qu

ence

EDIT

PRI

N

T

RU

N

fas36.eps

2. Press the

EDIT

soft key.

3. At the

Select Auto Sequence

prompt, press a numeric key (

1-9

), followed

by the

ENT

key. You are prompted to select, enable, or disable each of the

components listed below. Press the

ENT

key at each of the prompts to

advance to the next selection.

Print Auto Sequence Result

Pressure Gauge Test

Pressure Source Test

Pressure Leak Test

Pressure Relief Valve Test

Cuff Selection

NIBP Simulations (1-5)

Note

By default, each auto sequence is configured to perform the Leak Test,
Relief Test, and BP Simulation #1 (120/80) Changing the cuff selection
changes the available BP Simulation options.

4. When finished, either press the

RUN

soft key to run the sequence or press

HOME

to return to the

Main

menu.