Data sheet printout, Bp simulations – Fluke Biomedical BP Pump 2 User Manual

Page 63

Advertising
background image

Operation

Auto

Sequences

3

3-27

Data Sheet Printout

When this step is reached, a printout is generated, as shown in Figure 3-8, which
includes a standard header, the results of the previous tests, a place to record data
for the Pressure Source test (if enabled), and a list of BP Simulations that will be
executed.

BP Simulations

At the conclusion of the auto sequence, up to five different blood pressure
simulations are run in order of definition. Each simulation can be repeated as
many times as desired; however, each defined simulation must be run at least once
to advance to the next simulation.

Note

Some monitors cannot handle extreme changes in blood pressure. For
example, it may not be possible to perform a 255/195 simulation
following a 60/30 simulation on some monitors.