Introduction and Specifications

Instrument

Familiarity

1

1-9

Table 1-3. Number Key Functions

Number Name

Function

1

PRESSURE
LEAK

Pressurizes a pneumatic system to an
operator-defined target pressure up to
400 mmHg (53.3 kPa) and then
measures the loss of pressure over
time.

2

PRESSURE
RELIEF

Increases the pressure in the
pneumatic system until the relief valve
on the NIBP monitor opens or until the
Setpoint is reached, whichever occurs
first.

3

STATIC
PRESSURE

Accessed via the Pressure Gauge
Test, which enables the Tester to
measure static pressure generated by
an external source in the range of 50
to 400 mmHg 6.7 to 53.3 kPa).

4 STANDARD

BP

Provides seven variations of NIBP
simulations for both arm and wrist
cuffs.

5

PATIENT
CONDITIONS

Includes simulations for healthy,
geriatric, and obese patients, a well as
various levels of exercise.

6 ARRHYTHMIAS

Measures erratic heart rhythms,
including atrial fibrillation and
premature ventricular contraction.

7

RESPIRATORY
ARTIFACT

Exhibits a beat-to-beat variation in the
blood pressure caused by intrathoracic
pressure.

8 NEONATE

Tests the ability of the NIBP monitors
to detect blood pressure on neonatal
patients.

9

WRIST

Tests wrist cuff NIBP monitors.

0 USER

DEFINED

Allows the operator to define blood
pressure simulations.