Table D-1. Computer Control Commands (cont.)

Description Command

Returned

String

Perform Leak Test at
XXXmmHg

[LEAK,XXX]
50<=XXX<=400

Returns ACK
then Leak Rate.

Returns Leak
Rate after 60
seconds,
followed by CR-
LF, or NAK if
XXX out of
range•

Perform Pressure Source at
XXXmmHg

[PSOURCE,XXX]
50<=XXX<=400

ACK or NAK if
XXX out of
range

[BLEED_SYSTEM]

ACK

Performs Relief Valve Test

[RELIEF,XXX]
50<=XXX<=400

Returns ACK
then Peak
Pressure,
followed by CR-
LF, or NAK if
XXX out of
range

Pressure Gauge

[GAUGE] Returns

ACK

then pressure
port value,
followed by CR-
LF

Advertising