Pressure gauge, Leak test – Fluke Biomedical BP Pump 2 User Manual

Page 61

Advertising
background image

Operation

Auto

Sequences

3

3-25

Select A

u

to Se

qu

ence: 1

fas42.eps

3. Press a numeric key (

1-9

), followed by the

ENT

key.

The auto sequence begins to execute, based on the options that were enabled
using the

EDIT

function.

Note

Tests and simulations not enabled in the executing auto sequence are
skipped.

Enabled auto sequences are executed in the following order:

Pressure Gauge Test

Pressure Leak Test

Pressure Relief Valve Test

Pressure Source Test

Data Sheet Printout

NIBP Simulations

Pressure Gauge

This test monitors system pressure, which must be generated external to the Tester.

Press the

NEXT>

key to advance to the next test. Only the pressure displayed

when the

NEXT>

key is pressed appears on the printout.

Leak Test

This test performs a system leak test at the Setpoint pressure, which was defined
when the auto sequence was edited.

1. Press the

START

soft key to initiate this test. The system then pressurizes.

Once a stable pressure reading near the Setpoint is reached, a 60-second timer
begins to count down. Once the timer has elapsed, a Leak Rate appears on the
display.