Hold the restoration by the IPS e.max CAD Crystallization Pin.

Spray the IPS e.max CAD Crystall./Glaze Spray directly on the unfired

IPS e.max CAD Crystall./Shades and Stains. Spray the restoration from all sides

while simultaneously rotating it.

Shake the spray can again between individual bursts.

Spray an even layer onto the restoration.

Please observe the following procedure for the application of the IPS e.max CAD Crystall./Glaze Spray:

– Hold the restoration by the IPS e.max CAD Crystallization Pin. Do not spray the restoration while it is placed on the

IPS e.max CAD Crystallization Tray, as it would also be coated with glaze.

– Shake the spray can well immediately before use until the mixing ball in the container is moving freely

(approximately 20 seconds). If the spray is not sufficiently shaken, mainly the propellant is discharged

with a spraying burst. This, in turn, results in insufficient application of the glazing powder.

– Observe a distance of approx. 10 cm between the nozzle and the surface to be sprayed.

– Hold the spray can as upright as possible during spraying.

– Spray the restoration from all sides with short bursts while simultaneously rotating the restoration so that

an even, covering layer is created. Shake the spray can again between individual bursts.

– Spray the restoration a second time from all sides with short bursts while simultaneously rotating the

restoration so that an even, covering layer is created. Shake the spray can again between individual

bursts.

– Wait briefly until the glaze layer is dry and has assumed a whitish colour.

– Areas that do not show an even layer have to be sprayed again.

– Place the restoration in the center of the IPS e.max CAD Crystallization Tray.

– Conduct the Combination firing (Crystallization/Glaze LT/HT) using the stipulated parameters.