Firing parameters Crystallization/Glaze HT/LT

Please note:

– Crystallization/Glaze HT/LT can only be used for IPS e.max CAD HT and LT.

– If only max. two restorations are crystallized at the same time, the closing time S can be reduced to 1:30 min.

Furnace

P300

P500

P700

Stand-by

temperature

B

[°C/°F]

403/757

Closing time

S

[min]

6:00

Heating rate

t

1

[°C/°F/min]

90/162

Firing

temperature

T

1

[°C/°F]

820/1508

Holding time

H

1

[min]

0:10

Heating rate

t

2

[°C/°F/min]

30/54

Firing

temperature

T

2

[°C/°F]

840/1544

Holding time

H

2

[min]

7:00

Vacuum 1

1

1

[°C/°F]

1

2

[°C/°F]

550/820

1022/1508

Vacuum 2

2

1

[°C/°F]

2

2

[°C/°F]

820/840

1508/1540

Long-term

cooling

L

[°C/°F]

700/1292

Cooling rate

t

[°C/°F/min]

0

Note

Cooling after firing

– Remove IPS e.max CAD objects from the furnace after completion of the firing cycle (wait for the acoustic signal of the

furnace).

– Allow the objects to cool to room temperature in a place protected from draft.

– Do not touch the hot objects with metal tongs.

– Do not blast or quench the objects.