The following points should be observed for ceramic furnaces, used for the crystallization of IPS e.max CAD:

– Crystallization should be carried out in an Ivoclar Vivadent ceramic furnace (e.g. Programat P300, P500, or P700).

– If you would like to use other, untested ceramic furnaces, please consult Ivoclar Vivadent about their compatibility with

IPS e.max CAD.

– Basically, the following applies:

Ceramic furnaces without

– function for controlled (long-term) cooling

– vacuum function

cannot be used.

– Before the first crystallization and every six months after that, the ceramic furnace must be calibrated.

– Depending on the mode of operation, more frequent calibration should be required. Follow the instructions for the

manufacturer.

The following aspects should be observed for conducting the crystallization:

– Use only IPS Object Fix Putty or Flow as an auxiliary firing paste to place the restoration directly on the IPS e.max CAD

Crystallization Tray.

– IPS e.max CAD restorations must not be directly placed on the IPS e.max CAD Crystallization Tray and Pins, e.g. without

auxiliary firing paste, for crystallization.

– Use only the IPS e.max CAD Crystallization Tray and the corresponding Pins, since they store the heat necessary for slow

and above all tension-free cooling of the glass-ceramic.

– Always conduct the crystallization under vacuum.

– Remove IPS e.max CAD objects from the furnace after completion of the firing cycle (wait for the acoustic signal of the

furnace).

– Allow the objects to cool to room temperature in a place protected from draft.

– Do not touch the hot objects with metal tongs.

– Do not blast or quench the objects.

62

e.max

®

CAD

Crystallization Parameters

IP

S

Furnace

P80

P100
P200

P300
P500
P700

PX1

EP 600

EP 3000
EP 5000

Closing time

S

min.

6:00

6:00

6:00

6:00

6:00

6:00

Stand-by

temperature

B

°C/°F

403/757

403/757

403/757

403/757

403/757

403/757

Heating rate

t

1

°C/°F/min

60/108

60/108

60/108

60/108

60/108

Firing

temperature

T

1

°C/°F

770/1418

770/1418

770/1418

770/1418

770/1418

Holding time

H

1

min.

0:10

0:10

0:10

0:10

0:10

Heating rate

t

2

°C/°F/min

30/54

30/54

30/54

30/54

30/54

30/54

Firing

temperature

T

2

°C/°F

850/1562

850/1562

850/1562

850/1562

850/1562

850/1562

Holding time

H

2

min.

10:00

10:00

10:00

10:00

10:00

10:00

Long-term

cooling

L

°C/°F

700/1292

700/1292

700/1292

775/1427

1:30 min

700/1292

0:20 min

700/1292

700/1292

Cooling rate

t

l

°C/°F/min

0

0

Vacuum 1

1

1

1

2

°C/°F

550/1022
770/1418

550/1022
770/1418

550/1022
770/1418

550/1022
770/1418

550/1022
770/1418

Vacuum 2

2

1

2

2

°C/°F

550/1022
850/1562

770/1418
850/1562

770/1418
850/1562

770/1418
850/1562

770/1418
850/1562

770/1418
850/1562

Crystallization of IPS e.max CAD MO

Only the program memories 65-69 can be used with the Programat P100.

Please note:

– IPS e.max CAD HT and LT may also be crystallized with this program.

– If only max. two restorations are crystallized at the same time, the closing time S can be reduced to 1:30 min.