Precision Digital PD650 User Manual

Page 35

Advertising
background image

Model PD650 Large Display Process Meter

Instruction Manual

35

Set the first calibration point

Input 1 (INPT 1) flashes indicating that the meter
is ready to be programmed for the input for the
first calibration point. Press ENTER.

Inpt 1

Set the input for the first calibration point

The entire display will flash for three seconds. For
instructions see Five Basic Digit/Display Setting
Instructions
, page 30.

04.000

then

04.001

final input 1

04.003

Set the display for the first calibration point

Display 1 (Dspy 1) flashes indicating that the
meter is ready to be programmed for the display
for the first calibration point. Press ENTER.

Dspy 1

Program the display for display 1 (Dspy 1) as described
above when the input for the first calibration point was
programmed.

00.000

final display 1

02.500

To quit main menu, calibration, or scaling press ACK while displaying main menu
item or while display is flashing input n (InPt n) or display n (dspy n), where n is
the input or display number.

Minimum Input Span (Error Message)

The calibration input signals must be within the range of the meter and input 2
must be greater than input 1. However, the display for input 2 does not have to
be greater than the desired display for input 1.
If input 1 and input 2 signals are too close together an Error message will
appear and the display will return to the current input to be calibrated. To
terminate calibration process, press ACK button.

Input Range

Minimum Difference Between Input 1 & Input 2

0-5 V

0.16 V

0-10 V

0.32 V

4-20 mA

1.60 mA