Round horizontal tank calibration points, Number of points: 10 maximum error: 0.3% fs, Number of points: 8 maximum error: 0.5% fs – Precision Digital PD650 User Manual

Page 40: Calibration error ( error)

Advertising
background image

Model PD650 Large Display Process Meter

Instruction Manual

40

Round Horizontal Tank Calibration Points

Number of Points: 10 Maximum Error: 0.3% FS

Input mA

Display

(%

Volume)

1 4.00

0.00

2 4.80

1.80

3

6.00

7.20

4 7.20

14.20

5 9.20

28.10

6 14.80

71.80

7 16.80

85.80

8 18.00

92.80

9 19.20

98.10

10 20.00

100.00

Number of Points: 8 Maximum Error: 0.5% FS

Input mA

Display

(%

Volume)

1 4.00

0.00

2 4.80

1.90

3 6.40

9.50

4 8.40

22.40

5 15.60

77.70

6 17.60

90.50

7 19.20

98.10

8 20.00

100.00

Calibration Error (error)

A meter display of Error during calibration indicates that the calibration process
was not successful. The meter should be recalibrated.
The Error message will appear if input 1 signal and input 2 signal are too close
together. Refer to Minimum Input Span (Error Message), page 35.
The Error message will appear if input 1 signal is inadvertently also applied for
input 2 calibration, or ENTER is pressed before applying input 2.