Burkert Type 8056 User Manual

Page 17

Advertising
background image

- 17 -

RS232_485_ETP_MODBUS_BU_REV02.doc

3.2.15.

READING THE MAX-MIN ON INTERNAL DATA-LOGGER DATA OF ML210


The data logger can be read one record at a time with this command. If we assign the value AAH to the byte

specifying the address (byte 0), the data logger is cancelled. If we request the address of a record not present in the
memory, the returned data has no sense. For this purpose we recommend starting from the request for record 0 and

to check the number of records present in the memory which is returned by the measuring device..

Command code: 12
Command format: ADDRESS TO

+ ADDRESS FROM
+ 0CH

+ 01H

+ byte 0: index of the data logger data to read, AAH to cancel the entire content

+

CRC

CHECKSUM

Reply data format in the case where there are registrations present:

bytes 0-3:(32 bits IEEE floating point, MSB first) MAX FLOW RATE IN T.U.
bytes 4-7:(32 bits IEEE floating point, MSB first) MIN FLOW RATE IN T.U.

3.2.16.

LETTURA VALORI MIN-MAX DEL LOGGER INTERNO DI ML211

The data logger can be read one record at a time with this command. If we assign the value AAH to the byte

specifying the address (byte 0), the data logger is cancelled. If we request the address of a record not present in the
memory, the returned data has no sense. For this purpose we recommend starting from the request for record 0 and

to check the number of records present in the memory which is returned by the measuring device..

Command code: 12
Command format: ADDRESS TO

+ ADDRESS FROM
+ 0CH

+ 01H

+ byte 0: index of the data logger data to read, AAH to cancel the entire content

+

CRC

CHECKSUM


Reply data format in the case where there are registrations present :

bytes 0-3:(32 bits IEEE floating point, MSB first) MAX FLOW RATE IN T.U.

bytes 4-7:(32 bits IEEE floating point, MSB first) MIN FLOW RATE IN T.U.
bytes 8-11:(32 bits IEEE floating point, MSB first) MAX T. POWER IN T.U.

bytes 12-15:(32 bits IEEE floating point, MSB first) MIN T. POWER IN T.U.
bytes 16-19:(32 bits IEEE floating point, MSB first) deltaT MAX IN T.U.

bytes 20-23:(32 bits IEEE floating point, MSB first) deltaT MIN IN T.U.
bytes 24-27:(32 bits IEEE floating point, MSB first) T1 MAX IN T.U.

bytes 28-31:(32 bits IEEE floating point, MSB first) T1 MIN IN T.U.
bytes 32-35:(32 bits IEEE floating point, MSB first) T1 MAX IN T.U.

bytes 36-39:(32 bits IEEE floating point, MSB first) T1 MIN IN T.U.