Burkert Type 8056 User Manual

Page 58

Advertising
background image

- 58 -

RS232_485_ETP_MODBUS_BU_REV02.doc

Help command: A1CP2=?

Required access level: 2

min <> max

if accepted

1:CMD ERR if analog input not
enabled
5:ACCESS ERR if insufficient access

level

COMMANDS AND FUNCTIONS USED FOR READING THE PROCESS PARAMETERS

Name and description

Modes

Returned values or codes

Read command: FRSRN? Scale number: (1 or 2) if dual range

active

1:CMD ERR if dual range not enabled

Set command: NOT
SUPPORTED

1:CMD ERR if command is issued

FRSRN (Flow Rate Scale Range
Number) Reads the scale range

currently in use on the instrument.
Valid only if the dual-range
function is active.

Help command: NOT

SUPPORTED

1:CMD ERR if command is issued

Read command: FRVPC? Returns

%,value

string

Set command: FRVPC=

n

0:OK if parameters accepted

2:PARAM ERR if

n

out of range or

simulation mode not enabled

FRVPC (Flow Rate Value PerCent)

Reads or sets the flow rate value in
the percent form. Returns two
comma-separated fields, the

percent symbol and the numeric
value. Percent value is referred to

the full scale currently active. The
flow rate value can be set only
when the simulation mode is

active.

Help command:

FRVPC=?

min <> max (%)

Read command: FRVTU? Returns

measure-unit,value

string

Set command: NOT

SUPPORTED

1:CMD ERR if command is issued

FRVTU

(Flow

Rate

Value

Technical Units) Reads the flow

rate value and its technical
measure unit. Returns two comma-

separated fields, the technical unit
symbol and the numeric value.

Help command: NOT
SUPPORTED

1:CMD ERR if command is issued

Read command: VTTPV? Returns

measure-unit,value

string

Set command: NOT
SUPPORTED

1:CMD ERR if command is issued

VTTPV (Volume Totalizer Total

Positive Value) Reads the positive
total totalizer value and its

technical measure unit. Returns
two comma-separated fields, the
technical unit symbol and the

numeric value.

Help command: NOT

SUPPORTED

1:CMD ERR if command is issued

Read command: VTPPV? Returns

measure-unit,value

string

Set command: NOT

SUPPORTED

1:CMD ERR if command is issued

VTPPV (Volume Totalizer Partial

Positive Value) Reads the positive
partial totalizer value and its
technical measure unit. Returns

two comma-separated fields, the
technical unit symbol and the

numeric value.

Help command: NOT
SUPPORTED

1:CMD ERR if command is issued

Read command: VTTNV? Returns

measure-unit,value

string

Set command: NOT

SUPPORTED

1:CMD ERR if command is issued

VTTNV (Volume Totalizer Total
Negative

Value) Reads the

negative total totalizer value and
its technical measure unit. Returns

two comma-separated fields, the
technical unit symbol and the
numeric value.

Help command: NOT
SUPPORTED

1:CMD ERR if command is issued

Read command: VTPNV? Returns

measure-unit,value

string

VTPNV (Volume Totalizer Partial
Negative

Value) Reads the

negative partial totalizer value and

Set command: NOT
SUPPORTED

1:CMD ERR if command is issued

Advertising
This manual is related to the following products: