Instrument control, sample acquisition, session uploading, and
start, stop and pause features.

Pressurization

Normal air and sheath pressure readings vary between 6-9 psi while
the compressor runs. If the system pressure is out of range, your
sample acquisition will fail or return poor results.

Symptom

Possible Problem

Solution

PC cannot initialize
communication with analyzer.

The communications cable is
unplugged.

Check the communications
cable connections.

“Bad Link” error message

There is a bad link between the
PC, analyzer, and/or XYP.

Ensure that the instruments have
power.

Check the power cord and
communication cable
connections

Shut down the PC and analyzer.
Turn on the analyzer before
turning on the PC.

The Machine Setup is using the
wrong COM Port.

On the Analyzer menu, click
Setup. Verify that the correct
ports are selected for each
instrument. If you have make
changes, restart the system.

XY Platform microtiter plate
connect error. Command
timed out.

There is a bad link between the
XYP and PC.

Ensure that the communication
cable between the PC and XYP
is connected.

Symptom

Possible Problem

Solution

Pressurization Fails or
Pressure too low.

The sheath and waste lines are
not fully connected.

Make sure the lines between the
sheath and waste bottles and the
analyzer are fully connected.

An air leak is present in the
sheath bottle.

Remove and retighten the
sheath bottle cap.

The sheath or waste bottle
fittings are cracked.

Inspect the fittings to be sure
they form a tight seal.

There is a leak in the system.

Check for system leaks.