9 - 4

PN 89-00002-00-150 Rev. A

xMAP Microsphere
Stability and Storage

xMAP microspheres are light sensitive and should be protected from
light during all stages of storage and usage. In addition, freezing
conditions and organic solvents should be avoided.

Two-Step

Carbodiimide

Coupling of Protein

to xMAP

Carboxylated

Microspheres

Introduction

Use the protocols included here as a general starting point for
developing assays. All assays should be optimized for your reagents
in your specific application. You will discover the methods that give
you the best results by starting with these guidelines and modifying
them for your specific needs.

Equipment

Vortex

Sonicator bath

Micropipetters (1 µl - 1000 µl)

Microcentrifuge

Analytical balance

Timer

Rotator

Materials

xMAP carboxylated microspheres—LIMIT EXPOSURE TO
LIGHT!

Microcentrifuge tubes: 1.5 ml, polypropylene (See Technical
note 4)

ACTIVATION BUFFER:
0.1 M Monobasic Sodium Phosphate (pH 6.2 ± 0.2)

COUPLING BUFFER:
50 mM MES (pH 5.0 - 6.0)

WASH BUFFER:
Phosphate Buffered Saline (pH 7.4 ± 0.1), Tween

®

20

(0.05% v/v)

BLOCKING/STORAGE BUFFER:
Phosphate Buffered Saline (pH 7.4 ± 0.1), Bovine Serum
Albumin (10 mg/mL), Sodium Azide (0.05% w/v)

Advertising