Check fluid levels, Check for leaks, Calibrate – Luminex 200 User Manual with LDS 1.7 Software User Manual

Page 91: Sanitize

Advertising
background image

x

MAP Technology

Maintenance and Cleaning

PN 89-00002-00-150 Rev. A

7 - 3

Check Fluid Levels

Check the sheath and waste fluids daily, more often if you run
several samples. Replace the sheath fluid and empty the waste
container as needed. You must manually monitor the sheath fluid and
waste container levels. Release the Luminex analyzer’s pressure by
removing the lid from the sheath fluid container. Add sheath fluid,
filling the sheath fluid container to just below the air intake. Discard
waste by appropriate means. After you add sheath fluid and discard
the waste, remove air from the lines by clicking Prime until no more
air shows in the lines.

USE OF NON-LUMINEX APPROVED SHEATH FLUID
SHALL CONSTITUTE “IMPROPER USE” AND MAY VOID
WARRANTY RIGHTS PROVIDED BY LUMINEX AND/OR
ITS AUTHORIZED PARTNER.

Check for Leaks

‹

Check for leaks on a weekly basis.

Open all analyzer doors and visually inspect for leaks. If you see a
leak or spill, contact Luminex Corporation. Users in the U.S. and
Canada can call 1-877-785-BEAD (1-877-785-2323) between the
hours of 7:00 a.m. and 7:00 p.m. Central Time, Monday through
Friday. Users outside of the U.S. and Canada can call us at +1 512-
381-4397 between the hours of 7:00 a.m. and 7:00 p.m. Central
Time, Monday through Friday. Inquiries may also be sent by email to
[email protected].

Calibrate

Calibrate at least monthly during routine use. Follow the instructions
beginning on page 5-2.

Sanitize

‹

You should sanitize the system on a monthly basis and every

time you run biological samples.

1. Perform four Wash cycles using distilled water.

2. Dispense 1.2 mL of a fresh 10%-20% bleach

solution into a

sample tube. Place the tube into the sample tube holder, and
begin the Sanitize cycle.

Note:

In the event of a leak or spill

inside the analyzer, make sure the
power to the system is off and that
the power cord is disconnected
from the wall source.