Luminex 200 User Manual for LDS Version 1.7

x

MAP Technology

5 - 6

PN 89-00002-00-150 Rev. A

7. Fill the reservoir with a solution of 70% isopropanol or 70%

ethanol. Click Retract on the toolbar to retract the plate holder.

8. Click Alcohol Flush, then click OK. The alcohol flush removes

air from the fluidics. This takes about 5 minutes.

9. Click Eject on the toolbar and replace the alcohol in the

reservoir with distilled water. Click Retract on the toolbar to
retract the plate holder.

10. Click Calibrate on the main screen. The Start Calibration dialog

box opens.

Figure 5-7. Start Calibration Dialog Box

11. Enter your name in the Operator text box.

12. Select the Product / Lot Number for the xMAP classification

calibration and reporter microspheres.

13. Click OK. The Calibration dialog box opens.

Figure 5-8. Calibration Dialog Box

14. Make sure Classification is selected, and click Start.