3. Connect the new sheath filter, matching up the color-coded

fittings. The arrow on the sheath filter should point up.

4. Reconnect the sheath fluid bottle.

5. Prime the instrument 5 times to remove any air introduced into

the system.

Figure 7-4. Sheath Fluid Filter

Change Sheath

Quick-disconnect

O-rings

‹

Change the sheath quick-disconnect o-rings as necessary.

1. Turn off the Luminex analyzer at the power switch.

2. Remove the quick-disconnect fitting that has the faulty o-ring.

3. Use a small flat-head screwdriver to remove the worn out o-ring.

4. Place the new o-ring over the tip of a pair of closed needle-nose

pliers.

5. Gently separate the tips of the pliers to stretch the o-ring just

enough to fit it over the top of the quick-disconnect.

6. Slide the o-ring over the quick disconnect, then remove the

pliers.