Replace syringe seal – Luminex 200 User Manual with LDS 1.7 Software User Manual

Page 94

Advertising
background image

Luminex 200 User Manual for LDS Version 1.7

x

MAP Technology

7 - 6

PN 89-00002-00-150 Rev. A

Figure 7-2. Back of Luminex XYP, Removing Air Filter

3. Replace the filter and replace the screen.

4. Plug in and turn on the Luminex XYP.

Replace Syringe Seal

‹

Replace the syringe plunger seal every 6 months.

1. Turn off and unplug the Luminex analyzer.

2. Open the center door on the front of the Luminex analyzer.

3. Unscrew the knob on the syringe arm at the bottom of the

syringe, and forcefully push the syringe arm down.

4. Unscrew the syringe from the top of its housing; pull the plunger

out of the syringe.

5. Remove and replace the plunger seal.

6. Return the plunger to the syringe.

7. Screw the syringe back into the top of its housing.

8. Return the syringe arm to its original position, and hand-tighten

the screw on the syringe arm.

9. Plug in the power cord and turn the Luminex analyzer on.

10. Prime the Luminex analyzer 5 times and watch for any leaks in

the syringe area.

11. When finished, close the center door.

Warning:

The syringe arm does

not deactivate when changing
the plunger; injury could result if
the system is not unplugged.

Advertising