Roboinject

Setting Up the Roboinject

1. Place the Roboinject on a rigid, stable, and vibration-free base in an air conditioned room.

2. Provide a power supply in the immediate vicinity of the installation site.

3. Provide a compressed air supply in the immediate vicinity of the installation site.

4. Place the computer and its accessories next to it.

Connecting the Roboinject

The following illustration shows the rear panel of the Roboinject.

Connect all cables as described below.

Warning: Carefully lay and secure the cords. Remember that someone could easily trip over
a loose cable.

Note: All electrical connections are clearly marked, and the plug coding prevents confusion.
The cords should be plugged in without the use of excessive force.

Connecting the carrier

Plug the carrier connector into the according socket located on the Roboinject's rear panel,
labelled with "CARRIER OUT".

Warning: Confusion of the compressed air's inlet and outlet may destroy the device.
Have a close look at the preceding picture of the Roboinject's rear panel and take care
to connect the lines properly. The air hose connecting the Roboinject with the well plate
carrier is close to the electrical connection between them.