Roboinject

13

Connecting the Roboinject to the pressure line

1. Connect one end of the provided compressed air hose (ID 2.5 mm, OD 4 mm) to the water

separator.

2. Connect the other end to the compressed air inlet on the Roboinject's rear panel, labelled

with "AIR IN".

3. Connect the other provided compressed air hose (ID 4 mm, OD 6 mm) to the provided water

separator.

4. Make sure that the pressure line is closed. Connect the free end of the provided compressed

air hose (ID 4 mm, OD 6 mm) to the pressure line, labelled with "AIR OUT".

5. Open the pressure line and apply a pressure of 3 bar with the provided pressure regulator.

Warning: Make sure that the compressed air used fulfills all requirements for working with
the Roboinject. Otherwise, use of an improper compressed air could result in physical harm
to the hardware. Especially, check that the compressed air is absolutely free of water, oil,
and any particles. Refer to the chapter "

Important Safety Advice

" for more information.

Connecting the Roboinject to the computer

1. Connect the Roboinject by the provided USB cable, labelled with "USB".

2. Connect the Roboinject to the power line.

3. Finally, plug the power cord into the AC power line input, labelled with “24 VDC”.

4. Switch the Roboinject on or off via “I/O” toggle switch.

Warning: If the air cannot circulate freely around the external power supply, the device
may overheat. Do not shield the power supply by laying anything on top of it. Make sure
it is not exposed to direct sunlight.