Important Information and Instructions

2.1 Important Safety Advice

Warning: Make sure to read the following advice prior to installations of the Roboinject.
If you do not fulfill all requirements stated below, this may lead to malfunctions, breakage,
or even fatal injuries. Obey always the rules of local regulations and laws. Only qualified
personnel should be allowed to perform laboratory work. Work according to good laboratory
practice to obtain best results and to minimize risks.

The product has been built to the state of the art and in accordance with recognized safety
engineering rules. The device may only

 be used for its intended purpose;

 be used when in a perfect condition.

Improper use could lead to serious, even fatal injuries to the user or third parties and damage
to the device itself or other material damage. Malfunctions which could impair safety should
be rectified immediately.

2.1.1 High

Voltage

Electrical cords must be properly laid and installed. The length and quality of the cords must
be in accordance with local provisions.

Only qualified technicians may work on the electrical system. It is essential that the accident
prevention regulations and those of the employers' liability associations are observed.

 Each time before starting up, make sure that the mains supply agrees with the specifications

of the product.

 Check the power cord for damage each time the site is changed. Damaged power cords should

be replaced immediately and may never be reused.

 Check the leads for damage. Damaged leads should be replaced immediately and may never

be reused.

 Liquids may cause short circuits or other damage. Keep the power supply and the power cords

always dry. Do not handle it with wet hands.

2.1.2 Requirements for the Installation

 The Roboinject weighs more than 23 kg. Always grip it tightly and do not carry it alone,

but with the aid of another person.

 The movement of the well plate carrier can lead to vibrations of the workbench on which the

Roboinject is set up. Therefore, the Roboinject must be set up on a rigid, vibration-free base.
The base must also be sufficiently solid to carry the weight of the device.

 The Roboinject should be operated only in an air conditioned room. A room temperature of

20° C (or less) is recommended. Make sure that the device is not subject to direct sunlight.
It may overheat.

 If the air cannot circulate freely around the external power supply, the device may overheat.

Do not shield the power supply by laying anything on top of it.

 The external power supply unit is only for use with the Roboinject. Do not connect it to any

other instrument.