4.4 Setting Up the Well Plate

4.4.1 Inserting or Changing the Well Plate

Click the

icon to open the “Manual Mode” dialog. Select “Change Plate” in the “Global”

section to move the carrier into home position. Remove the used well plate if necessary. Place the
type of well plate which is defined in the “Plate Info” dialog in correct orientation on the carrier.

Place the well plate onto the carrier and press its side against the springs in the lower middle and
lower right corner of the carrier. Then put the rest of the plate down onto the carrier. The well
plate position is correct when there is no play between the well plate bottom and the carrier
surface at all edges and when the well plate has full contact to the left and upper side of the
carrier holding frame.

Important: The well plate should be immovably fixed on the carrier, otherwise the injection will
not be successful!