Roboinject

25

Click "Rinse" tab.

Rinsing the injection needle between the injection of different samples is mandatory for
preventing cross-contamination. In principal, you can reuse an injection needle as long as
it is not clogged or broken. Nevertheless, MCS does not recommend to reuse the needle
longer than one day after the last injection.

Define the number of "Repeats", the "Rinse Volume" in nl, and the "Velocity" in nl/s
from the up-down controls to avoid contamination when working with different compounds.
Volumes of 1000 nl and three repeats guarantee the absolute purity for each injected sample.

Click "Directories" in main menu "Settings". The "Set Directories" dialog opens.

Templates: Please browse through your folders, and define the directory you want to store
your template files with the extension "*.nit".

Plate Info: Please browse through your folders, and define the directory you want to store
your plate info files with the extension "*.rpf".

The "Log" file with the extension "*.log" is a file you can save by users command from
the "Log File" window. Please browse through your folders, and define the directory you
want to store the log files.