Multichannel Systems Roboinject Manual User Manual

Page 59

Advertising
background image

Injection

55

Enter an ID and / or a name which univocal identifies your sample. Select the type of “Sample”
and its “Concentration”.

Note: The minimal volume which always has to remain in the tube reservoir is 1 μl. For example,
if you need 4 μl sample volume for your injection run, please make sure that you start with at
least 5 μl. This is because the injection needle keeps always a safety distance to the tube bottom
to avoid uptake of contaminants and plugging.

Click the button "Set Injection" for changing injection parameters. The "Set Injection" dialog
opens. Please read also chapter "Inject" tab of "Options" dialog in main menu "Settings".

Ignore the "Set Injection" button for keeping the default settings.