3.9 Tube

View

In the "Tube View" you control the definition and selection of the samples. You can use
the eight reservoir tubes for eight different samples.

Click onto the symbol

of a "Edit Sample" icon.

The "Sample Information" dialog opens.

This dialog allows the definition of the respective injection samples inside of the 0.5 ml
micro reaction tubes.

Note: The sequence of injection is always from left sample (No 1) to right sample (No 8).
Please keep this in mind if the injection sequence of samples is of importance for you.

Enter an ID and / or a name which univocally identifies your sample in the "ID" and "Name"
window.

Select the type of sample RNA, DNA or Other in "Sample Type", and its "Concentration".

Important: The minimal volume which always has to remain in the tube reservoir is 1 μl.
For example, if you need 4 μl sample volume for your injection run, please make sure that
you start with at least 5 μl. This is because the injection needle keeps always a safe distance
to the tube bottom to avoid uptake of contaminants and clogging.

Enable the check box "Test Injection". The "Test Injection" dialog will appear before
the injection process of a sample is started. Please see chapter "Test Injection" dialog.

Type notes concerning the actual experiment into the "Comment" text box, for example,
the type of the sample when selecting "Others". The comments are saved in the "Plate Info"
file and shown in the report sheet.