Roboinject

41

Click the "Set Injection" button for setting injection parameters for that sample. Ignore the
button for keeping the default settings.

Click "OK" to complete the settings for this sample. Click "Clear" if you want to restart with
default values, and "Cancel" if you want to go back to the main window without changes.

Repeat the settings in "Sample Information" of other tubes until all samples you want to inject
are defined.

Note: The sequence of injection is always from left sample (No 1) to right sample (No 8). Please
keep this in mind if the injection sequence of samples is important for your injection process.

Click "Set Injection". The "Set Injection" dialog opens. The parameters are also available
in the "Inject" tab of the "Option" dialog in main menu "Settings". Please read chapter
"Inject" tab of the "Options" dialog in "Settings" menu.

After finishing the settings in the “Sample Information” dialog the tube icons appear colored,
according to their content for a good discrimination between the compounds.