Roboinject

37

Click "Next". The "Alignment" dialog of the "Start Injection Wizard" appears.

Plate Carrier: It is possible to move the plate carrier stepwise with the arrow buttons
in x- and y- direction on the table. The actual position of the carrier in relation to the home
position (X = 0 μm, Y = 0 μm) is displayed.

Use the arrow buttons to move the carrier stepwise in 30 μm steps (slow) or continuously (fast)
to any desired position. The X and Y values show the actual position relative to the home position
(X = 0, Y = 0).

Injector Mount: It is possible to move the injector mount stepwise with the arrow buttons
in z- direction on the plunger. The actual position of the carrier in relation to the home position
(Z = 0 μm) is displayed.

After proper alignment in x, y, and z- direction click "Finish".
The "Remove Microscope" dialog appears.

Please enable the check box if you do not want to see this dialog again. You can display
the dialog again when selecting the check box "Reset Remove Microscope dialog"
in the "General" tab of the options dialog.

Click "OK" and a sequence of movements are performed. First, the carrier moves to "waste"
position and the oil is extruded into the waste container. Then, the carrier moves to the first
defined sample tube and the necessary volume is taken up. Finally, the test injection dialog opens
if "Test Injection" was selected in the "Sample Information" dialog of the respective sample.